Risk | Low |
Patch available | YES |
Number of vulnerabilities | 1 |
CVE-ID | CVE-2021-43548 |
CWE-ID | CWE-20 |
Exploitation vector | Local network |
Public exploit | N/A |
Vulnerable software Subscribe |
Patient Information Center iX (PIC iX) Hardware solutions / Medical equipment |
Vendor | Philips |
Security Bulletin
This security bulletin contains one low risk vulnerability.
EUVDB-ID: #VU58250
Risk: Low
CVSSv3.1: 5.7 [CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H/E:U/RL:O/RC:C]
CVE-ID: CVE-2021-43548
CWE-ID:
CWE-20 - Improper input validation
Exploit availability: No
DescriptionThe vulnerability allows a remote attacker to perform a denial of service (DoS) attack.
The vulnerability exists due to insufficient validation of user-supplied input. A remote attacker on the local network can pass specially crafted input to the application and perform a denial of service (DoS) attack.
MitigationInstall updates from vendor's website.
Vulnerable software versionsPatient Information Center iX (PIC iX): B.02 - C.03
CPE2.3http://ics-cert.us-cert.gov/advisories/icsma-21-322-02
Q & A
Can this vulnerability be exploited remotely?
Yes. This vulnerability can be exploited by a remote non-authenticated attacker via the local network (LAN).
Is there known malware, which exploits this vulnerability?
No. We are not aware of malware exploiting this vulnerability.