Vulnerability identifier: #VU56080
Vulnerability risk: Medium
CVSSv4.0: 2.7 [CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:L/SC:N/SI:N/SA:N/E:U/U:Green]
CVE-ID:
CWE-ID:
CWE-434
Exploitation vector: Network
Exploit availability: No
Vulnerable software:
SpaceCom2
Hardware solutions /
Medical equipment
Vendor: B. Braun Melsungen AG
Description
The vulnerability allows a remote attacker to compromise vulnerable system.
The vulnerability exists due to insufficient validation of file during file upload. A remote attacker can upload any files to the /tmp directory of the device through the webpage API.
Mitigation
Install updates from vendor's website.
Vulnerable software versions
SpaceCom2: before 012U000062
Can this vulnerability be exploited remotely?
Yes. This vulnerability can be exploited by a remote non-authenticated attacker via the Internet.
Is there known malware, which exploits this vulnerability?
No. We are not aware of malware exploiting this vulnerability.